High Precision Film Flow Meter
SF-1U·2U
Overview
Available high precision flow rate measurement with one-touch operation.
High Precision: +-0.5%, +-1.0%
Flow Rate Measurement by utilizing a soap film detection.
LDC six digital display.
Remote Control by PC.
CE Mark Compliment
Specifications
SF-1U·2U |
 |
Display : |
LDC six digital display (Display can be toggled between;measured flow rate, measured time,temperature,and setting pressure.) |
Automatic measurement functions : |
Automatic measurement(2-10 times) and averaging are performed in AUTO mode (previously recorded data can be) |
Operating temperature range : |
5-40°C Guaranteed precision range:15-35°C |
External input/output : |
RS-232C compatible interface |
Power supply : |
AC adapter (7V DC) |
 |
VP-1U |
 |
Flow rate measurement range: |
0.2-10mL/min |
Precision: |
0.5%(±1%)0.2-2mL/min(2-10mL/min) |
Measured gases: |
N2, Air, O2, H2, He, Ar, CO, CH4, C3H8, etc
(Note:Water soluble gases cannot be measured, and setting pressure) |
Operating temperature range : |
5-40°C Guaranteed precision range(15-35°C) |
External input/output : |
RS-232C compatible interface |
 |
VP-2U |
 |
Flow rate measurement range: |
2-100mL/min |
Precision: |
0.5%(±1%)2-20mL/min(20-100mL/min) |
Measured gases: |
N2, Air, O2, H2, He, Ar, CO, CH4, C3H8, etc
(Note:Water soluble gases cannot be measured, and setting pressure) |
Operating temperature range : |
5-40°CGuaranteed precision range(15-35°C) |
External input/output : |
RS-232C compatible interface |
 |
VP-3U |
 |
Flow rate measurement range: |
20-1000mL/min |
Precision: |
0.5%(±1%)20-200mL/min(200-1000mL/min) |
Measured gases: |
N2, Air, O2, H2, He, Ar, CO, CH4, C3H8, etc
(Note:Water soluble gases cannot be measured, and setting pressure) |
Operating temperature range : |
5-40°C Guaranteed precision range(15-35°C) |
External input/output : |
RS-232C compatible interface |
 |
VP-4U |
 |
Flow rate measurement range: |
0.2-10mL/min |
Precision: |
0.5%(±1%)0.2-1L/min(1-10L/min) |
Measured gases: |
N2, Air, O2, H2, He, Ar, CO, CH4, C3H8, etc
(Note:Water soluble gases cannot be measured, and setting pressure) |
Operating temperature range : |
5-40°C Guaranteed precision range(15-35°C) |
External input/output : |
RS-232C compatible interface |
|
 |
|


|